Specifications
- Overall Height: 12′
- Chamber Diameter: 10′
- Chamber Interior Volume: 825 cubic ft
- Gross Weight: 24K lbs
- Nominal Operating Pressure: 30 psig
- Maximum Allowable Pressure: 33 psig
- Fabrication Hydrostatic Pressure Tested: 48 psig (>150% nominal operating pressure)
Note: Built in accordance with ASME (American Society of Mechanical Engineers) standards and practices under the strict specifications of Section 8, Division 1
Infrastructure Requirements
- Electrical
- One Dedicated 120V, 20 Amp, NEMA 5-20R Receptacle
- One Switched 120V, 15 AMP NEMA 5-15R Receptacle for Interior Lighting
- One # 6 AWG Connection to Building Ground System
- Plumbing
- Connection for Chamber Drain
- Connection to Air Supply
- Chamber Exhaust Penetrations
- HVAC Maintain Ambient Treatment Room Temperatures at Approximately 68 Degrees
Misc Equipment
- Air Supply Not Less Than 90 psig. If Not Already Available, Several Standard Models Available at Local Home Improvement or Hardware Stores will Suffice.
- Foundation. Standard Concrete Preparation As Per Architect or EOT Documentation
- Lightning Protection. As Per Local Contractor or Ordinance Mandate
- Standard NFPA Measures. As Per Local Fire Ordinance Where Required
Note: As Per Local and City Ordinances, Site Electrical and Plumbing Connections Must Be Performed By Local Licensed Contractors and are Therefore the Responsibility of the Site Builder